Determine the reference unit:
Press for about 4 seconds, until
[U. SEt.] is displayed, then
release
Change reference quantity:
Use the TARE-key or the arrow keys to
switch between the following reference quan-
tities:
10
5
100
Important: The larger the reference quantity,
the more accurate the parts counting.
Weigh in the reference unit:
Place as many parts to count on the balance
as the set reference quantity requires.
Reference optimization:
By adding more pieces (up to the 3-fold
quantity), you can optimize the reference. At
every reference optimization, the reference
weight is calculated anew. As the additional
pieces increase the base for the calculation,
the reference also becomes more exact.
46
or
30
The display shows flashing the last
saved reference quantity.
Qn 10
M
The display e.g. 10 Pcs. prompts
you to enter 10 pieces as reference.
Qn 30
M
The reference quantity is displayed
flashing.
30
The balance offers the possibility of
reference optimizing. If this is not
desired, press the F key.