7.2 Menu overview

The manufacturer's setting has a certain standard configuration. This one is marked
with *.
Function
Weighing mode
Additional functions
Zero balancing
Vibration filter
Display speed
Interface
(see chpt. 7.2.1)
Adjustment
* 1: Factory setting FEJ
* 2: Factory setting FES
FES/FEJ-BA-e-1132
Display
Selectio
or
or
* 1 Weighing
q SET.
* 0 Off
r SEL
s A.O
* 1 Automatic zero balancing is activated.
* 2
t S.D.
u RE.
* 3
* 1 6-digit data format
v K.F.
*1
w CA.
*2
Description of the
options
2 Combination: Weighing/parts counting
3 Combination: Weighing/percent determina-
tion
1 Add
2 Tolerance weighing
3 Combination Tolerance
weighing/adding
0 No zero balancing
Sensitive and fast (very quiet set-up location).
3
Robust but slow (very busy set-up location)
4
0 Setting for dispensing: Sensitive and fast
1
2
0 Deactivated
2 7-digit data format
3 extended 7-digit data
format
0 CAL-key turned off
1 Automatic internal adjustment (only FEJ)
2 Adjustment test with internal weight
(only FEJ)
3 External adjustment
4 Adjustment test with external weight
see chapter
12.3, Tab. 1
Insensitive but slow
see chap. 15.4.1
not documented