8.3.2 Default setting: Adjustment with internal weight

Condition: Function „i tCAL" activated (see ch. 8.3.1)
Press the [CAL] key The display shows "i-CAL"
Press the [TARE] key. The display changes automatically in this
order: "CAL 2", "CAL 1", "CAL 0", and "CAL End". After success-
ful adjustment the balance automatically returns to weighing
mode.
In case of an adjustment error (e.g. objects on the weighing
plate) the display will show an error message, repeat adjustment.

8.3.3 Default setting: Adjustment test with internal weight

During adjustment tests the balance automatically compares the saved value of the
adjustment weight with the actual value. This is only a check, i.e. no values are
changed.
Condition: Function „i tESt" activated (see ch. 8.3.1)
Press the [CAL] key The display shows "i-tESt"
Press the [TARE] key. The display changes automatically in this
order: "tESt 2", "tESt 1", "tESt 0"
After this, the difference to the previous adjustment will be dis-
played for several seconds.
After "tESt End" was displayed, the balance will automatically
return to weighing mode.
ABT-BA-e-1312
25