6.9.2 Adjustment test with external weight (only FES)

During adjustment tests the balance automatically compares the saved value of the
adjustment weight with the actual value. This is only a check, i.e. no values are
changed.
Procedure:
Observe stable environmental conditions. A warming up time (see chap. 1) is re-
quired for stabilization. Ensure that there are no objects on the weighing plate.
Operation
Activate function [w
(see chap. 7).
Start of the adjustment test:
Carefully place adjusting weight in the
centre of the weighing plate
The difference between the saved value
and the measured value is displayed.
FES/FEJ-BA-e-1132
CA. 4]
Display
w CA 4
a
CAL EH T
Qn
If the display indicates [ PuSH. F ], press
the
-key
Qn
If the display indicates [ PuSH. F ], press
the
-key
Qn
z0
g
CAL
K
CAL
O
CAL
F.5.
CAL
F.5.
23