Reading Water Quality - SciCan STATIM 7000 Operator's Manual

Cassette autoclave
Hide thumbs Also See for STATIM 7000:
Table of Contents

Advertisement

STATIM 7000 Operator's Manual
5. Maintenance
The factory default setting is for the unit to work with a water filter installed. If you choose
to use steam process distilled water instead of tap water, you can install the Water ByPass
Cartridge (SCWFBP) and set up the unit to work accordingly.
To set up the unit to work with the Water ByPass Cartridge, follow these steps:
1.
Remove the water reservoir and water filter as described previously and replace with
the Water ByPass Cartridge.
2.
Enter the User Menu (see section 3.4) and select the WATER FILTER option when it is
displayed on the first line.
3.
Using the
4.
Exit the user menu by pressing
To use the SciCan Water Filter Cartridge, you will need to enter the user menu and choose
INSTALLED under the WATER FILTER option.

5.8 Reading Water Quality

1.
Access User Menu. User menu can be accessed by either holding the
turning on the machine or by simultaneously pressing
is a idle and LCD displays "Select a Cycle"
2.
Using the
button.
>Water Quality
μS / yyy / z.z ppm
CD=XX
(down) &
(up) buttons, choose NOT INSTALLED, then press
and
buttons, scroll to Water Quality and select it by pressing the
CD = conductivity
yyy = engineering value
.
xx = micro Siemens value
z.z = parts per million value
26
5. Maintenance
button while
and
button while the unit
.