SciCan STATIM 7000 Operator's Manual page 17

Cassette autoclave
Hide thumbs Also See for STATIM 7000:
Table of Contents

Advertisement

4. Instructions For Use
4. Instructions For Use
2. Wrapped Cycles
The Wrapped Cycle is used to sterilize up to 3.4kg (7.5lbs) of solid and hollow metal
instruments which have been sealed or wrapped in autoclavable bags or sterilization wraps.
Instruments such as pliers, re-usable burs, scalers, forceps, dental and phaco handpieces
may be processed in this cycle. Consult the Instrument Mass Guide in the previous section for
information regarding approximate instrument weights in order to not exceed the maximum
load. At the end of the cycle's sterilization phase, drying will commence for 12 minutes.
Air-drying can be interrupted at any time by pressing the
be adjusted via the User Menu options.
To select the wrapped Cycle: Press the
to scroll through the available cycles.
WRAPPED 134°C / 3.5 min
WRAPPED 134°C / 18 min
Once the desired cycle is displayed, press the
button.
Please note that the 18 minute cycle is only available on some EN 13060 compliant models.
Additionally, this cycle must be unlocked by your Service Technician during setup as the
factory default is set to only provide the 3.5 minute cycle. Once the cycle has been activated,
the last unwrapped cycle that was run will be displayed the next time this cycle is selected.
This cycle is not available in North America.
button
(Not available in
North America.)
17
STATIM 7000 Operator's Manual
button. This drying duration can
A
B
C
A
100˚C / 101 kPa
B
6 purges (Conditioning)
peak at 133˚C / 295 kPa
vent to 115˚C / 169 kPa
C
Pressurizing
D
E
3.5
min /
18 min
D
Sterilizing
134˚C / 304 kPa -
138˚C / 341 kPa
E
Drying