STATIM 7000 Operator's Manual
3. Installation
3.4 Accessing the User Menu
The user menu allows the operator to reset warning messages, set the time and date, select
languages, and modify other items that do not affect the sterilization process of the unit.
The User Menu can be accessed by using one of two methods:
1.
Powering up the unit while pressing the
2.
Pressing the
"SELECT A CYCLE" message
3.5 Installing the Water or Pass-Through Filter
When the unit is first installed, either the provided water filter or the optional purchased pass-
through filter needs to be installed. Please refer to section 5.7 of this operators manual,
"Changing the Water Filter"
3.6 Filling the Statim Reservoir
The reservoir can be filled in two ways:
1.
By removing the reservoir from the unit and filling it seperately and then placing the
reservoir back into the unit.
2.
By removing the reservoir cover and pouring water directly into the reservoir.
If using the SciCan water filter (SCWF1), the reservoir can be filled using your local
tap water.
If not using the SciCan water filter, a Water Bypass Cartridge (SCWFBP) must be purchased
and used and then only use steam process distilled water containing less than 5 ppm total
dissolved solids (having a conductivity of less than 10 uS / cm).
The water conductivity can be checked on the LCD display by following procedure in Section
"Reading the water quality".
or
and
buttons simultaneously while the unit displays the
= Next option
= Previous option
= Select The Option
= Save & Exit or Return To The Main Menu
button