STATIM 7000 Operator's Manual
4. Instructions For Use

4.4 Running a Cycle

To operate each cycle, follow these steps and watch the LCD.
1. Turn the power switch at the back of the
unit to ON and wait for the LCD to display:
2. Choose the appropriate cycle by following
"Selecting a cycle" in section 4.3
3. Prior to the cycle starting, the unit's cycle
counter displays the total number of cycles
the unit has run to date.
Cycle you selected
CYCLE NUMBER
Cycle you selected
PRESS START
4. As the unit operates, the cycle progress is displayed on the LCD:
Cycle you selected
WARMING UP
CONDITIONING
110°C
As the cycle is running,
various sounds will be heard.
This is the normal functioning of the unit.
000000
143 kPa
PRESSURIZING
125°C
232 kPa
14:23
15/11/2007
SELECT A CYCLE
STERILIZING
136.8°C 330 kPa 3:29
20
4. Instructions For Use
VENTING
108°C
134 kPa