Advertisement

5. Maintenance
5. Maintenance
4.
Gently press the replacement filter into the filter bracket.
The arrow mark of the filter should be facing out and
pointing to the left. Re-attach tubes.
5.
Replace the water reservoir.
6 ..
Reset the warning message. To reset the
warning message, enter the User Menu and
select the AIR FILTER option when it is
displayed on the first line.
7.
Press the
on the first line.
8.
Press the
CYCLE screen.

5.5 Replacing the Cassette Seals

To ensure optimum performance of the Statim 7000, the seal is equipped with an electronic
RFID tag to ensure only authentic seals are used and also to provide the user with a LCD
message to replace the seal after 1000 cycles. The seal and O-rings should be changed every
1000 cycles or every 12 months.
Replacement seals are available from SciCan (p/n 01-110295S Cassette, Seal Kit).
When the cassette seal has undergone 1000 cycles, an early message on the LCD will display
SEAL LIFE WARNING - 100 CYCLES REMAINING, with a warning beep. The seals should be
changed immediately. After 1100 cycles, your unit will stop operating and a message will be
displayed SEAL LIFE EXPIRED, REPLACE. Replacement seals are available from your SciCan
dealer.
To change the cassette lid seal, follow these steps:
1.
Place the cassette lid and the new seal on a clean work surface.
2.
Examine the position of the old seal in the cassette lid and
arrange the new seal in the same orientation, next to the lid.
3.
Remove the old seal and discard.
4.
Clean any residue out of the seal channel and flush out the
channel with distilled water.
5.
Lubricate the new seal with the liquid seal lubricant provided.
6 ..
Align the RFID tag in the new seal with the holes in the lid.
7.
Insert the rounded edge of the seal under the round lip of the lid.
NOTE: When inserting the seal, seven round nibs should be visible. The nibs should fit flush
with the lid's outer surface. Ensure the seal is completely inserted. Feel around the periphery
to ensure the seal is securely seated.
button when RESET is displayed
button to exit and return to the SELECT
23
STATIM 7000 Operator's Manual
Air Filter
Water
Filter
RFID Tag
Figure 6
Figure 7