Maintaining Fluid Levels; Changing The Water Filter - SciCan STATIM 7000 Operator's Manual

Cassette autoclave
Hide thumbs Also See for STATIM 7000:
Table of Contents

Advertisement

5. Maintenance
5. Maintenance

5.6 Maintaining Fluid Levels

Unless you are using the SciCan Water Filter (SCWF1 Water Filter Cartridge), that allows the
use of tap water, you should only use steam-process distilled water containing less than 5
ppm total dissolved solids (having conductivity of less than 10 μS / cm) in the Statim. Each
time you refill the reservoir, empty the waste bottle. Empty the waste bottle often to avoid
unpleasant odors and discoloration of the contents. (A low-level chlorine-free disinfectant,
prepared according to the manufacturer's instructions, may be added to the waste bottle to
remedy this situation). The condenser bottle should be cleaned and refilled to the MAX line on
a weekly basis.

5.7 Changing The Water Filter

If you are using the SciCan Water Filter, it will need to be changed
regularly to ensure clean water is entering the internal components of
the unit. When the water filter cartridge is close to being exhausted,
a message will be displayed when a cycle is selected and an audible
tone reminds you to order or replace the filter. This message will appear
for approximately two seconds and then disappear, and then the chosen
cycle will start. This message will continue to appear every time a cycle
is chosen until the water filter is replaced with a new filter. If the filter is
not replaced and has been 100% used, the unit will not operate until a
new filter is installed.The water filter has a shelf life and should be used
prior to the expiry date inscribed on each filter.
When installed, the water filter has a maximum life of 6 .0 days, depending
on the quality of the tap water. The unit will notify you prior to the expiry
or exhaustion of the filter.
The water filter can be found under the removable water reservoir.
To replace this filter, follow these steps:
1.
Press the
2.
Remove the water reservoir.
3.
Remove the used water filter and discard.
4.
Remove the water filter from its packaging.
5.
Insert the new water filter. The water filter should be placed flat on
the bottom, with the tabs on the water filter facing upwards, and
then pushed inward until the filter is fully seated along the bottom
of the unit.
6 ..
Replace the water reservoir.
At the beginning of the next cycle, the unit will prepare the new filter for about 2 minutes and
display the message PLEASE WAIT.
button twice.
25