4. Instructions For Use
4. Instructions For Use
The buzzing noise during the air drying stage is the compressor operating. The air drying
phase of the cycle may be interrupted at any time by pressing the
will read:
When the air drying stage is finished the display reads:
If a sterilization cycle is successful the reminder tone will sound and the amber light flashes
until the
button is pressed or the cassette is removed from the unit.
Caution: Once the cycle has been completed, the cassette maintains aseptic conditions until
it has been opened.

4.5 Stopping a Cycle

To stop a cycle press the
or the unit detects a problem while operating, the cycle will stop. Once a cycle has been
stopped, the
button must be pressed before another cycle can be started. The display
reads any of the following messages:
If the display shows the message, CYCLE FAULT or NOT STERILE, the cassette contents are
not sterile! See section 7. Troubleshooting, for more information.
If the air drying stage of the cycle is interrupted, do not store wrapped instruments that were
in the cassette unless they are dry.
PLEASE WAIT
CYCLE COMPLETE
REMOVE CASSETTE
CYCLE COMPLETE
button. If the
button is pushed, the cassette is removed,
CYCLE FAULT XXX
NOT STERILE
or
CASSETTE REMOVED
NOT STERILE
21
STATIM 7000 Operator's Manual
button. The display
STOP BUTTON PRESSED
CHECK FOR DRYNESS

Hide quick links: