Alarm Code A040 & E041 Automatic Water Filling Problem - SciCan Bravo Troubleshooting Manual

Hide thumbs Also See for Bravo:
Table of Contents

Advertisement

Error A040 – Filling Problem Alarm
Failure to automatically fill the Bravo sterilizer from the external water tank, the MIN level signaling is
not turned OFF within 2 minutes from the start command of the automatic filling.
Note: If Bravo is a new installation or was completely drained you may need to turn the Bravo OFF
and back ON to allow the unit to automatically fill twice to get sufficient water to activate the MIN
level switch.
1. Check for water in the external filling tank.
2. Check for a kinked tubing from the external filling tank to the Bravo sterilizer.
3. Check that the valve on the external filling tank is open.
4. Check water filter is not clogged or cracked. If the water filter is cracked the pump may be louder
than normal when running.
5. Check the auxiliary pump is providing water to the reservoir. Pump is located in rear of Bravo
mounted vertically.
6. Prime Auxiliary Water Pump
7. Disassemble and clean Auxiliary Pump
Error E041 – Filling Problem
In automatic fill mode unit is filling too frequently. Two automatic fills every two cycles.
1. Check automatic fill bottle for water.
2. Check tubing from automatic fill bottle to back of unit for kinks.
3. Check water filter on input of Auxiliary Pump.
4. Check Auxiliary Pump.
1/4" Tubing
Basting Syringe
Bravo Error A040 & E041
Filling Problem Alarm
Auxiliary Pump Priming Procedure