STATIM 7000 Operator's Manual
5. Maintenance
To change the cassette tray seals (O-Rings), follow these steps:
1.
Remove the perforated rack (01-110294S)
2.
Remove the tray exhaust duct (01-110297S)
3.
Unscrew the exhaust coupling (Cassette Outlet Coupling 7000, 01-110292S) and
carefully remove the O-ring and seal.
4.
Find the matching seal from the kit and install onto the port.
5.
Secure the exhaust coupling to the cassette ensuring the metal spring is on the outer
wall of the cassette.
6.
Unscrew the inlet coupling and steam deflector (Cassette Inlet Coupling 7000,
01-110291S) and carefully remove the O-Ring and seal.
7.
Find the matching seal from the kit and install onto the port.
8.
Secure the second coupling to the cassette ensuring the metal spring is on the outer
wall of the cassette.
9.
Replace the exhaust
duct and perforated
rack.
NOTE: During a cycle, steam
may appear between the lid
and the tray. If this persists,
remove the cassette and
check that the seal is
correctly installed.
Be careful. The metal parts
may be hot, and the cassette
may contain hot steam.
For the unit to function
effectively, the steam
deflector and perforated
rack MUST be placed
back in the cassette in
the proper position.
Perforated Rack
Inlet O-Ring
Steam Deflector & Coupling
Exhaust Coupling
Figure 8
24
5. Maintenance
Inlet Spring
Inlet Seal
Inlet Coupling Nut
Exhaust
Coupling Nut
Exhaust Spring