7. Troubleshooting
7. Troubleshooting
Message WATER QUALITY
IS NOT ACCEPTABLE.
Machine will not start.
Message REFILL
RESERVOIR, Machine will
not start.
Time and date are incorrect.
Message MSD NOT
CONNECTED <> INSERT
MSD/FLASH
Message MSD/FLASH FULL
<> REPLACE MSD
If you are not using the water filter, you may have used
water that is not steam-process distilled or is improperly
distilled.
Empty the reservoir and refill with steam-process distilled
water containing less than 5 ppm total dissolved solids
(having conductivity of less than 10 μS / cm). If you have
the water conductivity meter, using the user menu, check
the quality of the water before refilling the reservoir.
The level of the water in the reservoir is low. Refill the
reservoir.
The time and date have not been set. Set the time and date.
Check the serial cable connection.
Check the power connection.
Ensure the lower red LED is lit.
Check that the Mass Storage Device is properly inserted.
Repeat the instructions for Installing the SciCan Data
Logger on your Statim.
The MSD is full. Export the data.