Maintaining Fluid Levels; Reading Water Quality; Using On-Screen Instructions - SciCan statim 2000 G4 Operator's Manual

Cassette autoclave
Hide thumbs Also See for statim 2000 G4:
Table of Contents

Advertisement

9. Maintaining your STATIM

9..7 Maintaining Fluid Levels

1. The water reservoir level is continually monitored by your STATIM. If the reservoir is low,
a red X will appear on the water icon
to the next screen to confirm that it is a water level issue and not a water quality issue.
2. If the reservoir is low, a
3. To fill the reservoir, remove the cap from the top of the unit and fill the reservoir.
We recommend using a funnel to minimize spills. Each time you refill the reservoir,
empty the waste bottle and refill with water to the MIN line. Empty the waste bottle often
to avoid unpleasant odors and discoloration of the contents. (A low-level chlorine-free
disinfectant, prepared according to the manufacturer's instructions, may be added
to the waste bottle to remedy this situation).

9..8 Reading Water Quality

1. The water quality is continually monitored by your STATIM. Use only steam-process
distilled water containing less than 5 ppm total dissolved solids (having conductivity of
less than 10 μS / cm) in the STATIM. If the water quality is beyond these tolerances,
a red X will appear on the water icon
to go to the next screen to confirm that it is a water quality issue and not a water
level issue.
2. If the water quality not suitable,
3. Using the drain tube (see section 3.5 Priming the Pump, Figure 6) empty the contents
of the reservoir into the water container and replace with steam-process distilled water
containing less than 5 ppm total dissolved solids (having conductivity of less than
10 μS / cm).

9..9. Using On-Screen Instructions

The STATIM G4 on-screen help instruction can be viewed by following these steps:
1.
2. Scroll to
3. Choose the Instructions you would like to view and select.
will appear next to WATER LEVEL.
Instructions
and select.
Page 44
of the cycle select screen. Press the icon to go
of the cycle select screen. Press the icon
will appear next to micro S. and parts per million values.