13. Appendices
CT Mode
Applied Tube Voltage Peak [kVp]
Applied Tube Current [mA]
Direction [°]
0
45
90
135
180
225
270
315
Test Condition
Tested Mode
Mode
Ultra Low Dose, FOV 100 x 80 5.9s [mR/hr]
1 m (3.3 ft)
588.2
549.3
472.6
458.8
12.9
410.5
663.2
429.7
138 I
PaX-i3D Smart
CT FOV 100 x 85 18s Low Dose
99
14
1.5 m (4.9 ft)
135.3
249.4
307.3
287.6
4.6
288.7
301.4
194.2
2 m (6.6 ft)
87.1
106.8
78.4
89.3
1.3
98.2
112.4
92.3