10.
Cleaning and Maintenance
10.2.2

Maintenance Task Checklis

Before operation, ensure that the equipment is clean and ready for
use. Make sure that all parts which come into contact with the patient
have been cleaned.
After using the equipment, make sure that the main power switch has
been turned off.
Ensure that the equipment is firmly plugged into a dedicated power
source.
Ensure that the plug and power cord are not hot.
Confirm that the orange (exposure) indicator lamp turns on when
the exposure switch is pressed. Ensure that the orange (exposure)
indicator light remains on for the entire duration of the exposure.
Ensure that the power cable is not kinked, broken, exposed and that it
is free of all other defects.
Confirm that activating the emergency stop switch ceases the unit's
operation. Pressing the emergency stop switch should stop all
equipment movements and X-ray emission.
Ensure that all visible labels are intact and legible.
Check for possible wear or damage to the exposure switch cable.
Confirm that the audio message is audible throughout the duration of
the exposure.
Maintenance Tasks
Please refer to the "Constancy Test Requirement" for documents about the image
quality inspection methods.
116 I
PaX-i3D Smart
Maintenance
period
Daily
Daily
Daily
Daily
Daily
Daily
Weekly
Weekly
Monthly