X-Ray Scatter Dose - Vatech PaX-i3D Smart User Manual

Hide thumbs Also See for PaX-i3D Smart:
Table of Contents

Advertisement

13. Appendices
13.2.2

X-ray Scatter Dose

Instrument
Radiation Monitor
Controller
PANO Mode
Mode Tested
Applied Tube Voltage Peak [kVp]
Applied Tube Current [mA]
Mode
Direction [°]
0
45
90
135
180
225
270
315
Test Equipment Information
Manufacturer
Model
Radcal
9015
Mode Tested: PANO Normal Adult 13.6s
PANO Normal Adult 13.6s [mR/hr]
1 m (3.3 ft)
60.9
19.6
10.6
22.1
1
45.4
47.6
76.4
137 I
PaX-i3D Smart
S/N
91-1470/19069
PANO HD
80
14
1.5 m (4.9 ft)
2 m (6.6 ft)
17.7
12.4
6.8
12.5
0
21.4
21.9
19.4
8
5.8
4.1
5.6
0
9.4
9.2
8.6