Low Temperature Threshold (deg. C)
Format: <K-degrees,message>
degrees:
message:
0 to 50, zero dis-
Any 1 to 7 character ASCII string except NUL, <, or >.
ables.
Default is LO-TEMP.
default is
50
Example: To change the temperature to 45 and message to LO-TMP,
send <K+,LO-TMP>.
Lifetime Hours
Format: <K$hours10,message>
hours10:
message:
1 to 65,535
Any 1 to 7 character ASCII string except NUL, <, or >.
default is
2500
Default is
Hours entered are 1/10 of the actual hours.
Example: To change the Lifetime Threshold to 20,000 and change message to MAINT,
send <K$2000,MAINT>.
Present Operating Temperature (deg. C)
Format: <K%degrees>
degrees:
0 to 125
Read Only. Send
<K%?>
to read.
Counts
Format: <K_powerups,resets>
powerups:
resets:
0 to 65,535
0 to 65,535
Read Only. Send
<K_?>
to read.
Powerups are the number of times the power is turned off and on. Resets include
Watchdog Timeout, <A>, <Z>, <Zd>, and hardware defaults.