Assembly Replacement Procedures
A5 Solid State Drive
A4A1 CPU Memory Card
1. Refer to
Figure 15-37
CPU Memory Card Removal
Disk Drive Interconnect
Removal
1. Remove the A5 Disk Drive assembly for the instrument. Refer to the
2. Remove the A4 CPU assembly from the instrument. Refer to the
3. Refer to
Keysight NFA Series Noise Figure Analyzers Service Guide
Figure
15-37. Remove the SSD as described on
memory card (2) can be removed by pushing on the memory card and it
will spring out. To install the memory card push it into the slot until you
feel the spring catch. Then replace the SSD as described on
Solid State Drive"
removal procedure.
Assembly"
removal procedure.
Figure
15-38. Remove the A4 CPU assembly cover (1) by first
removing the 19 flat-head screws (2) and 3 pan-head screws (3). The 19
flat-head screws need to be discarded and replaced since they are
precoated with a thread locking compound and new screws will need to be
used when the cover is re-installed.
page
435. The
page
436.
"A5
"CPU
437