Functional Tests
Amplitude Accuracy at 50 MHz
Figure 17-4
Amplitude Accuracy Test Setup

Procedure

1. Zero and calibrate the power meter.
2. Configure equipment as shown in
To minimize stress on the test equipment connections, support the power
sensor.
3. If the auto alignment for the analyzer has not been performed within the
4. Press MODE/MEAS. Select the Spectrum Analyzer mode, Swept SA
5. Set up the synthesized sweeper by pressing:
6. Set up the signal analyzer by pressing:
502
connected directly to the signal analyzer input through the adapter.
past 24 hours, press System, Alignments, Align Now, Align Now All to
perform the auto alignment routine.
measurement, and Normal view. Press OK. Press Mode Preset.
CW, 50, MHz
Power Level, -4, dBm
RF (On)
Input/Output, Freq Ref Input, Freq Ref Input External
FREQ, Center Frequency, 50, MHz
Span, 2, MHz
Figure
17-4, with the power splitter
Keysight NFA Series Noise Figure Analyzers Service Guide