Assembly Replacement Procedures
Input Connector Assembly
Replacement
1. Refer to
Make sure that the O-ring
2. Reconnect the semi-rigid cable W10 to the Input Connector Assembly.
3. Place the gusset (5) in position and replace the three screws (4) that
4. Replace the Front Frame Assembly. Refer to the
5. Replace the instrument outer case. Refer to the
456
Figure
15-50. Place the Input Connector Assembly (2) into
position in the chassis. Replace the two screws (1) to attach the Input
Connector Assembly to the chassis. Torque to 9 inch-pounds.
Torque to 10 inch-pounds.
attach it to the chassis. Torque to 9 inch-pounds.
replacement procedure.
replacement procedure.
is installed correctly on the input connector.