Testing Preamp Option (P03, P07, P26, P44) - Keysight N8973B Service Manual

Nfa series
Hide thumbs Also See for N8973B:
Table of Contents

Advertisement

Functional Tests
Amplitude Accuracy at 50 MHz

Testing Preamp Option (P03, P07, P26, P44)

Instruments containing Options P03, P07, P26, P44 must have the preamp
function turned on and tested.
Procedure
1. On the analyzer, press MODE/MEAS, Spectrum Analyzer, OK, AMPTD,
2. Connect the equipment as shown in
3. Set the synthesized sweeper amplitude to the value listed in the Nominal
4. Set the signal analyzer input attenuation to 0 dB.
5. Set the Span as listed in
6. Record the source amplitude, as measured by the power meter, in the
7. On the signal analyzer, press Restart.
8. Wait for the analyzer to finish averaging.
9. Press Peak Search.
10.Record the signal amplitude as measured by the analyzer in the measured
11.Calculate the signal amplitude accuracy using the following equation:
12.Record the results under the Amplitude Accuracy Error column of
Table 17-7
Amplitude Accuracy Results (Option P03, P07, P26, P44)
Nominal
Source
Low- band
Amplitude
Preamp
(dBm)
-13
ON
504
Signal Path, Internal Preamp, Low Band.
30 dB Pad between the adapter and the signal analyzer input.
Source Amplitude column in
Power Meter Amplitude column of
amplitude column of
Table
Amplitude Accuracy Error = Meas_amp + 30 dB - Power_meter
17-7
Res BW
Span
(kHz)
(kHz)
1
106
Figure 17-4 on page
Table
17-7.
Table
17-7.
Table
17-7.
17-7.
Measured
Power Meter
Amplitude
Amplitude
Meas_amp
Power_meter
(dBm)
(dBm)
Keysight NFA Series Noise Figure Analyzers Service Guide
502, using a
Table
Amplitude
Test Limit
Accuracy
(dB)
Error
(dB)
±1.30 dB

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N8974bN8975bN8976b