Sd Memory Card Removal Procedure - Keysight X Series Manual

Signal analyzers
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

7

SD Memory Card Removal Procedure

When present, the SD Memory Card is located on the A4 Processor assembly.
However, it can only be accessed after the instrument's disk drive is removed.
To remove the SD Memory Card:
1. Remove the disk drive by following the
2. Push in on the SD Memory Card and release. The Card springs back out,
page
25.
Once the disk drive has been removed, the SD Memory Card is accessible.
and can then be removed.
instrument rear panel, with the Disk Drive Assembly and SD Memory Card
extracted from the instrument.
Security Features and Document of Volatility
SD Memory Card Removal Procedure
"Disk Drive Removal Procedure" on
Figure 7-1 on page 29
below shows the
28