Functional Tests

Test Equipment

Test Equipment
The table below summarizes the test equipment needed to perform all of the
functional tests. Alternate equipment model numbers are given in case the
recommended equipment is not available. If neither the recommended nor the
alternative test equipment are available, substitute equipment that meets or
exceeds the critical specifications listed.
Table 17-1
Analyzer
Model
Item
Adapters
N8973B,
3.5 mm (f) to 3.5 mm (f)
74B, 75B
(connector saver for source)
N8976B
2.4 mm (f) to 2.4 mm (f)
N8976B
2.4 mm (f) to 3.5 mm (f)
(connector saver for source)
All
BNC (f) to SMA (m)
All
Type N (f) to Type N (f)
All
Type N (m) to 3.5 mm (m)
All
Type N (m) to 3.5 mm (f)
All
Type N (f) to 3.5 mm (f)
All
Type N (m) to BNC (f)
Attenuators
All
10 dB Step Attenuator
All
10 dB Fixed Attenuator
(2 required)
Keysight NFA Series Noise Figure Analyzers Service Guide
Critical Specifications
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 44 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
Frequency: 40 MHz
Frequency: 10 MHz to 18 GHz
VSWR: < 1.05:1
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 1 GHz
VSWR: < 1.05:1
Frequency: 50 MHz
Range: 0 to 70 dB
Frequency: 50 MHz
VSWR: < 1.2:1
Al ternate
Recommended
Keysight
Keysight Model
Model
1250-1749
83059B
11900B
11901B
1250-1200
1250-1472
1250-0777
1250-1743
1250-1744
1250-1745
1250-1476
8495A
8496A
8493C Option 010
8493A Option
010
or
8493B Option
010
489