Further Calibration Information; Difference Between Full-Range And User-Range Calibration - Omicron Lab Bode 100 User Manual

Hide thumbs Also See for Bode 100:
Table of Contents

Advertisement

Calibrating Open/Short/Load:
For more details and practical examples on how to perform calibration in the Voltage/Current
measurement mode, please refer to the corresponding application notes on
www.omicron-lab.com.
8.4

Further calibration information

8.4.1 Difference between Full-Range and User-Range calibration

Full-Range calibration and User-Range calibration differ only in the frequencies that are used to
measure the correction factors.
Full-Range calibration measures the correction factors over the "full" frequency range of the
instrument at factory-predefined frequencies.
User-Range calibration measures the correction factors at exactly the same frequency range and
frequency points that are used in the measurement currently configured by the user.
OMICRON Lab
Open calibration.
Short calibration.
Note that the inductance of the short connection
is assumed to be zero. Short Delay Time is 0 s by
default.
Load calibration.
Default Load Resistor value setting is 50 Ω.
Calibration
87