Result format equations
The Impedance / Admittance / Reflection results are calculated based on the following equations:
ImpedanceZ:
where R is the real part of the impedance (resistance) and X the imaginary part of the
impedance (reactance).
ω is the angular frequency
Admittance Y:
where G is the real part of the admittance (conductance) and B the imaginary part of
the admittance (susceptance).
ω is the angular frequency
40
with the frequency f.
Series equivalent resistance Rs equals the real part of impedance
Series equivalent capacitance Cs
Series equivalent inductance Ls
Series circuit quality factor Q
tan(d) of impedance
with the frequency f.
Parallel equivalent resistance Rp
Parallel equivalent capacitance Cp
Parallel equivalent inductance Lp
Parallel circuit quality factor Q
tan(d) of admittance