Measurement types and applications
7.4

Impedance analysis

7.4.1 One-Port

The One-Port measurement mode is designed to perform quick impedance, admittance and reflection
measurements. One advantage of the One-Port impedance measurement mode is that no external
calibration is required. Bode 100 can measure impedance directly at the output port.
Measurement information
Bode 100 derives the impedance by evaluating the internal source voltage and the output voltage.
Receiver 1 as well as Receiver 2 are internally connected. General details on impedance
measurements with Bode 100 can be found in
7.2 Impedance measurement introduction
on page 37.