Performing A Gain Calibration - Omicron Lab Bode 100 User Manual

Hide thumbs Also See for Bode 100:
Table of Contents

Advertisement

8.2

Performing a Gain Calibration

Gain calibration (also called Thru calibration) is used to remove the gain and phase error introduced
by the connections between the measurement instrument and the DUT in a gain or transmission
measurement.
To achieve maximum accuracy, do not change the attenuators after having performed an
external calibration.
Hint: Thru calibration can also be used in impedance measurements that are based on a
gain measurement. These measurement modes are e.g. Shunt-Thru, Series-Thru or Voltage/
Current. When performing a Thru calibration in such a mode, the underlying gain
measurement is calibrated and afterwards the calibrated gain result is transformed to
impedance.
Performing the calibration
In order to calibrate a gain measurement, please proceed as follows:
1. Build up the calibration setup. For details on the calibration setup, please see
Transmission (S21) measurement
measurement
on page 73.
2. In the Bode Analyzer Suite, click on the Gain calibration icon (either User-Range or Full-Range).
To learn more about the difference between User-Range and Full-Range calibration please check
8.3.1 Difference between Full-Range and User-Range calibration
3. The calibration dialog opens and the calibration state shows Not Performed.
4. Ensure that the calibration setup is connected properly and press Start.
5. Wait until the calibration has completed and the calibration state shows Performed.
6. Close the calibration dialog.
70
on page 71 or
8.2.2 Calibrating a Gain/Phase
8.2.1 Calibrating a
on page 87.
OMICRON Lab

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents