Calibrating A Shunt-Thru Or Series-Thru Measurement - Omicron Lab Bode 100 User Manual

Hide thumbs Also See for Bode 100:
Table of Contents

Advertisement

8.3.4 Calibrating a Shunt-Thru or Series-Thru measurement

In this section you learn how to calibrate an Impedance, Reflection or Admittance measurement in the
Shunt-Thru or Series-Thru measurement mode.
Shunt-Thru and Series-Thru are based on a S21 Transmission measurement. Bode 100 measures
S21 and the Bode Analyzer Suite calculates impedance from the S21 measurement. Details on the
calculation can be found in
You can either calibrate the underlying S21 measurement using Thru calibration or you can use Open/
Short/Load calibration to move the reference plane directly to the DUT.
Shunt-Thru and Series-Thru allow both, Thru or Open/Short/Load calibration.
However, only one calibration can be active at a time!
Even if both calibrations have been performed only one calibration is applied.
You must select the calibration you want to apply using the slider in the calibration dialog.
If the arrow points to the left, Thru calibration is applied.
If the arrow points to the right, Open/Short/Load calibration is applied (see example below).
Shunt-Thru calibration connections
Calibrating Thru:
Calibrating Open/Short/Load:
84
7.2.3 Shunt-Thru
on page 64 and
7.2.5 Series-Thru
Thru calibration can e.g. remove the gain and
phase error introduced by the connection cables
including a coaxial common mode transformer
that is generally used in this measurement to
suppress the cable-braid error.
Open calibration.
on page 66.
OMICRON Lab

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents