In the Advanced Settings region you can change the following values:
1. The Load Resistor value represents the value of the resistor used for Load calibration.
2. Short Delay Time is the time delay of the Short element used for Short calibration.
The default settings for Load Resistor and Short Delay Time depend on the
measurement mode.
If you change a value from its default a warning sign will be shown
Changing Load Resistor or Short Delay Time will delete the corresponding calibration!
OMICRON Lab
Calibration