Section 3: HART commissioning
April 2014
3.4.2

Device dashboard menu tree

Figure 3-1. 3144P Device Dashboard HART 5- Overview
3144P HART Revision 5
Menu Trees - Overview
Key
Bold = Menu Title
Underline = Method
Overview
Configure
Service Tools
32
Overview
1. Device Status
2. Primary Variable
3. Sensor 1 Temp
4. Sensor 2 Temp
5. Analog Output Value
6. Upper Range Value
7. Lower Range Value
8. Device Information
Active Alerts
1. Good (only if no alerts)
2. Failed (Fix Now)
3. Advisory
4. Maintenance (Fix Soon)
Device Information
1. Identification
2. Revisions
3. Alarm Type and Security
Reference Manual
00809-0100-4021, Rev GC
Good
1. No Active Alerts
Failed
1. Invalid Configuration
2. Electronics Failure
3. Calibration Error
4. Sensor Drift Alert Active
5. Sensor 1 Open
6. Sensor 2 Open
7. Sensor 1 Shorted
8. Sensor 2 Shorted
9. Terminal Temperature
Failure
Advisory
1. Calibration Error
Maintenance
1. Hot Backup Active
2. Sensor Drift Alert Active
3. Sensor 1 Degraded
4. Sensor 2 Degraded
5. Excess EMF
6. Sensor 1 Out of Operating
Range
7. Sensor 2 Out of Operating
Range
8. Terminal Temperature Out
of Range
9. Primary Out of Limits
10. Analog Output Saturated
11. Analog Output Fixed
Identification
1. Tag
2. Model
3. Electronics S/N
4. Final Assembly Number
5. Date
6. Descriptor
7. Message
8. Device Image
Revisions
1. Universal
2. Field Device
3. Hardware
4. Software
5. DD Revision
Alarm Type and Security
1. Alarm Direction
2. High Sat Level
3. Low Sat Level
4. High Alarm Level
5. Low Alarm Level
6. Security Switch
HART® commissioning