To review the configuration of the thermocouple diagnostic: Fast Keys
2-2-4
1.
2.
3.
4.
5.
HART® commissioning
From the Home screen, 3 Service Tools.
Select 4 Maintenance.
Select 1 T/C Diag Snsr 1 or 2 T/C Diag Snsr 2 depending on which sensor you are
interested in.
Select 3 TC Diag Config to view the configuration information of your sensor.
To Reset Baseline Value: If you wish to reset the baseline value of your sensor, select 4
Reset Baseline and press OK.
Section 3: HART commissioning
April 2014
111