Specifications - Canon CXDI-50C Service Manual

Cxdi-1 system
Table of Contents

Advertisement

4
S
p
e
c
i
f
i
c
a
t
i
4
S
p
e
c
i
f
i
c
a
t
i
The CXDI-50G/50C (Imaging Unit/Power Box) is the Digital Cassette that has the mobility and
can be used on the optional angles.
(1) Imaging Unit
This unit consists of the internal sensor, 50Di Board, 50AD Board, 50LED Board and its
outer cover. The sensor unit converts the X-ray image to the electrical signal (O/E
Conversion) and after performs the A/D conversion, transfer its signal through the Power
Box with Ethernet cable to the Control PC.
Item
Object
Effective filming range
Number of Pixels
Effective Number of Pixels
Pixel pitch
Fluorescent substance
Output gradations
Transfer method
Imaging cycle
Dimension
Imaging Unit coloring
Imaging Unit mass
(except Grid)
Space between surface where
patient gets in contact (CFRP)
and sensor surface (glass)
Remote SW OFF
Heat
Sleep
emission
Max load mode
Remote SW OFF
Power
Sleep
Consumpt
-ion
Max load mode
*
Mechanical strength
Control PC
Power Control
(ON/OFF)
Grid attach/remove detector
Cable for Imaging Unit
Count of connected sensor
Scattered radiation backward
block sheet
Environment-conscious
unleaded type
Photo timer
o
n
s
o
n
s
50G
General Shooting Cassette
353 x 430mm
2214 x 2700
2208 x 2688
160µm x 160µm
GOS Fluorescent screen
12bit (4,096 gradations)
A/D 14bit
Ethernet: Imaging Unit to
Control PC
(Through the Power Box)
15 sec. (standard)
491 (W) x 477 (D) x 23 (H)
mm
Cool white
4.8Kg
5.7Kg
5.2Kg
Within 5.0±0.5mm.
9.5kcal/h
19kcal/h
24kcal/h
11W
22W
28W
Cassette with resisting the
strength
(Original specification)
CXDI-C3S
General PC in market
None
(Power Box: Operation
with Remote switch
manually)
Yes
3/7m 2types
Multiple Imaging Units are
connectable to a single
control PC via Network
switch. (Limitation of the
number of connectable
Imaging Unit depends on the
specification of control
software in control PC)
Mo sheet (0.3 mm thick)
Cannot be built in
1. General
50C
CsI
4.9kg
5.8kg
5.3kg
Within 4.9±0.5mm
30kcal/h
35W
5
Remarks
(Mobile/Desktop PC)
one panel
Including handle portion
427 mm (excluding handle portion)
Except the cable
With 7m cable
With 3m cable
Including the Power Box
Including the Power Box
Including the Power Box
1 image per 15 Sec.
Including the Power Box
Including the Power Box
Including the Power Box
1 image per 15 Sec.
Load uniformly: 150Kg
φ
Load partly: 100Kg/
40mm
The Imaging unit is put on the plain surface
with the Sensor side (Detector) is up.
Distinguish with Product order.
Network switch should be procured at each
sales company.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cxdi-50g

Table of Contents