Canon CXDI-50C Service Manual page 41

Cxdi-1 system
Table of Contents

Advertisement

3.2 Network setting for Screwcap.ini
The CXDI software communicates with the imaging part through screwcap.dll by using the
communication protocol for sending and receiving commands and responses.
In conjunction with the communication, Screwcap.ini retains the information of the connected
CXDI-50G/50C and control PC such as network addresses.
It is necessary to edit Screwcap.ini to communicate with the imaging part.
Since the Screwcap.ini factory setting is the following default setting (see the figure below), it is
not necessary to edit the setting unless you changed the network protocol TCP/IP setting for the
LAN card that communicates with the CXDI-50G/50C imaging part in the previous item, "3.1
TCP/IP setting for the control PC network card".
Screwcap.ini is located in the following directory:
D:\ccr\screwcap.ini
[HostInfo]
ResponseTimeout=30000
DataTimeout=60000
IntervalTimeout=30000
[SensorInfo]
SensorNum=1
[Sensor1]
IpAddress=192.168.100.11
CommandPort=12121
DataPort=12122
[Log]
LogLevel=2
1
*
CXDI-50G/50C's IP address:
2. Installation Manual
[SensorInfo]
SensorNum=1 Confirm this
CXDI-50G/50C IP address.
This must be the same as the
IP address stored in the
imaging part.
This address must be the same as the IP address stored in
the next item, "3.3 Network setting stored in the CXDI-
50G/50C sensor".
- 24 -
*1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cxdi-50g

Table of Contents