4. Repair Guide
(2) Attaching
Attachment of CXDI-50G/50C Grounding Strengthening Parts.
1. Requirements
(1) Grounding strengthening parts
• PCB-50LED (BG7-2767-100)
• Aluminum spacer (BA4-2434-000)
• Plain washer M2.6mm (XD1-2100-267)
• Spring washer M2.6mm (XD1-3100-267)
If procedures in Service Manual Report CXDI-50G 05-013 has already been conducted, the above plain
and spring washers should already be attached.
(2). Tools
• Torque screwdriver (recommended)
• Tester (for measuring resistance value)
• Cable attachment jig (BY9-6551)
• Adhesive
Cable attachment jig is used when attaching the new cable. For details, see Service Manual Report
CXDI-50G 05-001.
2. Notes
• Do not place anything in the chest pocket. Things in your pocket may fall out and damage the
sensor unit.
• Be sure that the work place is well-ordered to prevent the chassis from getting scratched.
• Be sure that the work place is protected against static electricity.
3. Disassembly
See the service manual for the procedure for removing each parts.
(1) Remove the rear cover from the sensor unit.
(2) Remove the sensor cable.
(3) Remove PCB-50LED. (Handle the FFC cable carefully.)
18