Canon CXDI-50C Service Manual page 67

Cxdi-1 system
Table of Contents

Advertisement

Parameter
-v
This parameter dumps the
transfer log.
-jn
This is the time to take
timeout.
-k
level=0
1
2
3
Argument: Values necessary for executing a function, subroutine, procedure, or other operation is
passed to them. Arguments are assigned to functions and subroutines when executed. For example,
the argument in f (x) is x.
PDU: Protocol data unit
The types of PDU's include get-request, get-next-request, get-response, set-request, nd trap.
Note
For details about the "-v" parameter, see "Checking the Error Log".
Note
The parameters "-v" and "-d" put the log on the console. Therefore, be sure to always erase
these parameters before operation by the user.
If –k:DICOM Modality LUT OD is enabled, set appropriate options to suit each output
destination.
level = 0: Do not delete (default when not specified)
1: Delete Window Center/Width
2: Delete Window Center/Width and Rescale Intercept/Slope/Type
3: Delete Rescale Intercept/Slope/Type (compatible with releases up to Ver.4.20)
Other than a loadable LUT or γ=1.0 has been specified with the output destination-
specific LUT function.
For storage: Specify -k3(or -k2).
If IMG Rescale Type = 0D is unidentifiable to storage, resulting in an error: Specify -k3.
(If only one storage is connected, simply set DICOM Modality LUT OD to "Disabled.")
The implementation allows Window Center/Width with Rescale Type = 0D specified to
be interpreted as "optical densityx1000."
For storage: Specify -k1.
The CXDI Window Center/ Width output value (implementation) is fixed at 2048/4096.)
Note
With AGFA impax Ver. 4.5.0, the specification of -k0 demonstrated a successful density-
intensity conversion. But because the corresponding text in the DICOM specifications
document is ambiguously written such that the status of implementation by other
manufacturers is unknown, please be advised to consult the storage manufacturer for each
connection destination or work out on a trial and error basis. Also note that an external
storage option specification (if (DICOM modality LUT OD is set to "Enabled," OD tags
<0028, 1052 - 1054> are assigned and Level: 0 is assumed) is not supported.
2. Installation Manual
Meaning
* This parameter is used
simply as "-v".
* DUL and SRV are dumped.
* This parameter is used to
make the transfer software
put the debugging character
string on the console.
* Sets the time to take timeout
in seconds.
* A variety of specifications
have since been needed in
conjunction with DICOM
modality LUT support.
- 50 -
Description
* This parameter does not affect DICOM
data transfer.
* This parameter is to be changed when
taking timeout.
If DICOM Modality LUT OD is enabled,
set appropriate options to suit each
output destination.