Canon CXDI-50C Service Manual page 82

Cxdi-1 system
Table of Contents

Advertisement

Precautions for connecting the server (reference)
Equipment
Transmitting the 0019 shadow group causes a failure, and the group is not received
Kodak
properly with the default. Its reception is enabled by setting the strictValidation
Miil
parameter to Off in Miil.
Fujitsu
Dr. ABLE
Hitachi
send_image error return values and log output (reference)
Example of
error
Successful
Return value: 0x00
No log output at this time
Invalid parameter
Return value: 0x00000001
exists.
CXDI interpretation:
See below for the log output at this time:
-------------------------------------------------------------------------------------
-t
-v
node
port
image
-------------------------------------------------------------------------------------
Server fails to start.
Return value: 0x00180012
CXDI interpretation:
See below for the log output at this time:
-------------------------------------------------------------------------------------
18-132600[127]ERR:
18-132600[127]ERR:
18-132600[127]ERR:
-------------------------------------------------------------------------------------
As a result of
Return value: 0x10
providing the
CXDI interpretation: DICOM response error (CCRTRANS_ERR_RESP)
server with a
See below for the log output at this time:
command request,
-------------------------------------------------------------------------------------
an error was
returned as
-------------------------------------------------------------------------------------
response.
2. Installation Manual
Restrictions on connections
Transmission is currently performed with the "1 study multi series/1 series 1
image" setting. However, since a multi format is used for the screen displays for
each series under the Dr.ABLE specifications, the switching operations are a
hassle. The user will find it more convenient if it is at all possible to change the
setting to "1 study 1 series/1 series multi image." (These unusual data
specifications were requested with the full understanding of their unusualness.)
With DICOM, the body parts (such as the abdomen and head) and their directions
(such as PA and AP) belong to the series information. It therefore follows that a
different series is required for a different body part or body part direction. This
aspect is restricted by the DICOM standard rather than by the installation and
other steps taken by us. To put it the other way around, multiple images with
different body parts and their directions cannot be put together as a series. To
remedy this problem, devising a way of enabling the viewers to reference
different series at the same time at some future point in time will be helpful.
Transmission is currently performed with the "1 study multi series/1 series 1
image" setting. However, since a multi format is used for the screen displays for
each series, the switching operations are a hassle. The user will find it more
convenient if it is at all possible to change the setting to "1 study 1 series/1 series
multi image."
This problem arises with the Fujitsu equipment as well. Refer to the section on
Fujitsu.
Return value and log output
(CCRTRANS_ERR_DICOMPARAM)
Set called AE title to title in Association RQ
Place DUL and SRV facilities in verbose mode
Node name for network connection
TCP / IP port number of server application
A list of one or more images to send
(CCRTRANS_ERR_NOT_CONNECT)
connected)
DICOM connect error
DICOM connection error
60012 TCP Initialization Error: Invalid argument
130012 Peer aborted Association (or never
180012 Failed to establish association
(To be determined)