Canon CXDI-50C Service Manual page 109

Cxdi-1 system
Table of Contents

Advertisement

2. Installation Manual
2) Look in the X-ray tube to check the cross line reflected from the mirror matches the cross line
on the X-ray tube. If it does not match, adjust the height of the table by the adjuster or the
spacer to align the slight differences.
Reflected cross line
Cross line of X-ray tube
Correct adjustment
Adjuster balance is wrong
Look in
Cast the cross line
Reflect the cross line
(4) Regardless the height of the X-ray tube and the sensor unit, it should keep the positional relation
aligned in step (3) above.
x
- 92 -