Removing Grip unit

2.1
Base cover
(1) Remove the handle cap with the handle cap hook in the handle cap hole pushing.
(2) Loosen the screw (x2) to remove the handle plate and handle part.
Note:
1. Handle with carefully so that the defect should not adhere.
2. Check the direction of the handle cap hole at the assembling. Adjust surface which opens
the hole to the Base cover side and fix.
3. The handle cap must be pushed tightly until the hook is in the handle cap hole position.
[Replacing Handle Unit]
To improve the performance of the handle unit, following parts have been changed.
Key No.
Description
HANDLE,
1
IMAGING UNIT
2
CAP, HANDLE
PLATE,
3
HANDLE
Former and new parts are not interchangeable.
Replace three parts from Key No.1 to 3 in sets to allow the compatibility with a sensor.
CXDI-50G
CXDI-50C
4. Repair Guide
a
s
s
e
m
b
l
e
a
s
s
e
m
b
l
e
Screw (x2)
Hook
Former Part No.
BA4-1792-030
BA4-1793-020
BA4-1794-020
7M
3M
7M
3M
4
Handle cap (x2)
New Part No.
Qty
BA4-1792-040
1
BA4-1793-030
2
BA4-2992-000
2
Body No. 150827 -
Body No. 250001 -
Body No. 100056 -
Body No. 200001 -
Handle cap hole
Handle plate (x2)
Remarks