4-3) Select the END STUDY and transfer the SMPTE pattern image to the printer or the
high-definition monitor. [Fig. 6]
4-4) On a printed film or on the monitor, measure the density of 32 steps of the grayscale
on the test image. [Fig. 7]
*1: Measure the SMPTE image density on the film using a densitometer. The SMPTE
image density on the high-definition monitor is measured using the gradation
analysis software.
*2: The data for the SMPTE test image grayscale transferred by the CXDI are the
values for the maximum density (3.20 in the case of the MLP 190) in the printer
settings which have been changed into 32 uniform steps.
2. Installation Manual
[Fig. 6]
[Fig. 7]
- 73 -
Step1
Step32