(14) Backing Up When Installing
1)
Purpose
In case of re-installing the CXDI application, the necessary files ex) the exposure
position and other parameters must be backed up so that can be restored at the status of
first installation.
2)
Necessary items
2-1) Removable drive such as MO drive or external HDD that can connect to USB 2.0
3)
Notes
3-1) Before performing backup procedure, delete any "BodyPart" and image data exposed
for tests.
Deleting image data: Refer to the "Deleting Data" item.
Deleting "BodyPart": Refer to the CXDI Series Operation Manual.
3-2) Backup should be performed immediately before handing over the product to the
customer only for new installations.
4)
Connections
4-1) Check that all equipment is turned off.
4-2) Connect the keyboard and mouse removal drive to the control PC.
5)
Settings
5-1) Turn on the operation unit power and then the control PC power.
5-2) The CXDI application starts. Press the [Alt] + [Tab] keys to switch the program to the
Command Prompt screen.
5-3) The message "Welcome to Canon CXDI" appears. Input [8] and press the [Enter] key.
(Select "8 Exit".)
5-4) The Windows XP desktop screen appears.
5-5) Right click on My Computer on the Desktop screen, and select Explorer from Menu.
5-6) Explorer appears. Copy the CCR folder onto Removable Drive.
2. Installation Manual