Canon CXDI-50C Service Manual page 91

Cxdi-1 system
Table of Contents

Advertisement

4-5) As the step (3)-6), create a graph based on the data measured in step 4), and make sure
that the data from the minimum density to the maximum density nearly form a straight
line.
Characteristics of the printer may prevent the data from forming a straight line near
the minimum and maximum densities. Rotate the image on the QA screen, and reprint
or retake measurements.
3.50
3.00
2.50
2.00
1.50
1.00
0.50
0.00
1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31
Ideal Density Data for Linear Line
Step
1
2
3.2 D
0
0.10
3.1 D
0
0.10
12 Bit
0
132
8 Bit
0
8
Step
12
13
3.2 D
1.14
1.24
3.1 D
1.10
1.20
12 Bit
1453
1585
8 Bit
90
98
Step
23
24
3.2 D
2.27
2.37
3.1 D
2.20
2.30
12 Bit
2906
3038
8 Bit
180
188
2. Installation Manual
Step
3
4
5
0.21
0.31
0.41
0.20
0.30
0.40
264
396
528
16
24
32
14
15
16
1.34
1.45
1.55
1.30
1.40
1.50
1717
1849
1981
106
114
122
25
26
27
2.48
2.58
2.68
2.40
2.50
2.60
3170
3302
3434
196
205
213
- 74 -
6
7
8
0.52
0.62
0.72
0.50
0.60
0.70
660
792
924
41
49
57
17
18
19
1.65
1.75
1.86
1.60
1.70
1.80
2113
2245
2377
131
139
147
28
29
30
2.79
2.89
2.99
2.70
2.80
2.90
3566
3698
3830
221
229
238
Max setting
density
Film
base
density
9
10
11
0.83
0.93
1.03
0.80
0.90
1.00
1056
1188
1320
65
74
82
20
21
22
1.96
2.06
2.17
1.90
2.00
2.10
2509
2641
2774
155
164
172
31
32
3.10
3.20
3.00
3.10
3962
4095
246
255

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cxdi-50g