Canon CXDI-50C Service Manual page 130

Cxdi-1 system
Table of Contents

Advertisement

3. Function
4. Imaging conditions table
It is possible to add values up to 3000 msec and 4800 mAs to the system settings -
imaging conditions table on the settings screen. However, if 1001 - 3000msec, 1601 -
4800mAs values are added, a warning message will be displayed only once, saying the
values cannot be used with the CXDI-11 - CXDI-31 sensors. [Figure 3-2]
If at the time of a regular exposure or a calibration exposure, the imaging body part of a
CXDI-11 - CXDI-31 sensor was selected, values higher than 1000 msec, 1600 mAs
cannot be selected from the imaging conditions table.
Press [OK]
Figure 3-2
11