Specifications - Canon OCT-A1 Operation Manual

Optical coherence tomography
Table of Contents

Advertisement

Appendix

Specifications

OCT Image
Mode
Resolution
Scanning speed
Angle of view
Light source
SLO Image
Scanning mode
Angle of view
Resolution
Light source
Capturing
Scan modes
• [Macula 3D]
• [Glaucoma 3D]
• [Disc 3D]
• [Wide 3D]
• [Custom 3D]
• [Multi Cross]
256
Optical spectrum OCT
Depth (Z direction): 3 μm (in-house data)
Width (X-Y direction): 20 µm
70,000 A-scan per second
34 degrees (H) × 34 degrees (V) (10 × 10 mm)
SLD 855 nm
Confocal laser scanning mode
45 degrees (H) × 34 degrees (V) (13 × 10 mm)
25 µm
LD 780 nm
1024 A-scan (H) × 128 B-scan
Scanning area: 10 × 10 mm
1024 A-scan (V) × 128 B-scan
Scanning area: 10 × 10 mm
512 A-scan (H) × 256 B-scan
Scanning area: 6 × 6 mm
512 A-scan (V) x 128 B-scan
1024 A-scan (H) x 1 B-scan
Scanning area: 13 x 10 mm
1024 A-scan (H/V) × 128 B-scan
Scanning area: 3 to 10 mm (adjustable range)
Horizontal: 1024 A-scan (H) × 5 B-scan
Vertical: 1024 A-scan (V) × 5 B-scan
Scanning area:
Horizontal: 3 to 13 mm (adjustable range)
Vertical: 3 to 10 mm (adjustable range)
Number of averaging: 1, 5, 10 sheets
Number of averaging (*): 1, 5, 10, 20, 30, 50 sheets