Parameter List (Separate Document 1)
DICOM storage device
In the CXDI, DICOM data transfer is performed using the transfer software module
"send_image" .The settings for these parameters are described below.
Parameter
-m maxPDU
Maximum PDU value in
byte units
-t calledTitle
Called App Entity Title
-c callingtitle
calling App Entity Title
-s SOPName
(for reference)
This parameter designates
whether class be connected
for performing association
at the beginning of
transfer.(CR/T/MR/NM/S
C/US)
-I
A-RELEASE-RES is
ignored.
-d FAC
This parameter dumps a
specific facility log.
(DCM/DUL/SRV)
2. Installation Manual
Meaning
* The CXDI automatically
uses 131072 internally for
operation.
* Designating a specific value
allows overwriting of the
above value.
* The AE Title setting field is
automatically applied to this
setting.
* Designating a specific value
allows overwriting of the
above value.
* The CXDI automatically
uses CANON_CCR
internally for the operation.
* Designating a specific value
allows overwriting of the
above value.
* This is not used in the
CXDI.
* This parameter is used
simply as "-I"
* This parameter is used
simply as "-d"
* This parameter is used to
make the transfer software
put the debugging character
string on the console.
- 49 -
Description
* The DICOM standards do not allow
values of 1301073 or higher to be set.
* This is used when the operator who
manages the connected storage device
requests a size change.
* In DICOM printing, note that the
argument title changes to -u.
(
See the printing parameters.)
* Note that the meaning is opposite of the
DICOM printing argument -c.
( See the printing parameters.)
* This is used when the operator who
manages the connected storage device
requests a change in the installed
identification information (version).
* Note that the meaning is opposite of the
DICOM printing argument -c.
( See the printing parameters.)
* This is used when the operator who
manages the connected storage device
requests a change in the installed
identification information (version).
* This is used when the error message
"30012 Peer aborted Association (or
never connected)" occurs even though
the DICOM data transfer was
successful.
This is used differently based on the
connected storage devices.
* This parameter does not affect DICOM
data transfer.
* This parameter outputs the CXDI log
based on Windows NT.