Maintenance Intervals

Always clean and disinfect the unit and/or its components
before maintenance and before sending it/them in for
repair.
O
sensor
2
Cooling-air filter
Lip seals of the connecting
plate behind patient system
NiCd battery for mains failure
alarm
Time-keeper RAM
Inspection and service
General overhaul of pressure
reduction valves
CAUTION
For units with LC displays:
If the glass of the LC display is broken, a chemical liquid can
leak out. Avoid skin contact.
If affected, wash skin areas immediately with soap and water!
replace in the case of a fault message
(page 17).
clean or replace every 4 weeks (page 94);
to be replaced at least once per year.
to be replaced at least every 2 years by
trained service personnel
to be replaced at least every 2 years by
trained service personnel (see page 94 for
disposal).
to be replaced at least every 4 years by
trained service personnel (see page 94 for
disposal).
to be carried out every 6 months by
trained service personnel
to be carried out every 6 years by trained
service personnel
Care

Maintenance Intervals

93