Care
Checking readiness for use
Release the hose and then disconnect the connector from
the Y-piece.
The red alarm lamp must blink and the alarm must sound.
Display (example):
Airway pressure low
or
Leak in hose system?
Check setting!
Set the rotary knob »PEEP/CPAP« back to 0.
Connect the test lung again.
If it passes all of the above tests the Babylog 8000 plus
is ready for use!
Test the humidifier as described in the related
Instructions for Use!