Before Using For The First Time - Dräger Babylog 8000 plus Instructions For Use Manual

Intensive care ventilator for neonates software 5.n
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Preparation

Before using for the first time

CAUTION
When an extractive monitor is uses, a vacuum may be
generated in the airway if the inspiration hose becomes
blocked; risk of injury to the patient!
For this reason:
Connect the sampling line of the extractive monitor only via
the adapter with safety valve 84 12 448. The Luer Lock
connector must be at the top in order to avoid the formation
of condensation.
Before using for the first time
The built-in battery for the power-failure alarm is charged
during normal operation of the ventilator. Before the ventilator
is used for the first time, or after a long idle period, it should be
switched on for 30 minutes in order to charge the battery
sufficiently.
Use the following settings in order to prevent alarms during this
charging period:
1
Set the rotary knob »O
2
2
Set the rotary knob »Insp. Flow
3
Set the rotary knobs »T
»T
4
Set the rotary knob »P
insp
5
Set the rotary knob »PEEP/CPAP« to 3.
6
Switch on the ventilator by pressing the power switch on the
rear until it locks.
CAUTION
If the main switch is not pressed until it locks into position, but
is only pushed under the flap, the device may be unintentio-
nally switched off. When switching off, the device generates
the power failure alarm.
22
-Vol%« to 21.
« to 5.
« to 0.4,
I
« to 0.6.
E
« to 20.
1
3
3
2
4
5
6