Reading The Log - Dräger Babylog 8000 plus Instructions For Use Manual

Intensive care ventilator for neonates software 5.n
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Operation

Reading the log

Suppressing the message and audible signal for about
30 seconds:
1
Press the key »OK«.
Each message is automatically recorded in the log of the
ventilator.
Suppressing the warning/alarm tone for 2 minutes
2
Press the key » «.
CAUTION
While the warning/alarm tone is suppressed for 2 minutes,
new messages will still be displayed,
but without the warning/alarm tone.
Switching the warning/alarm tones on again:
2
Press the key » « again.
A summary of the fault messages with their causes and
remedies, can be found on page 78 etc.
Reading the log
Each alarm, caution and advisory message is written into the
electronic log of the ventilator.
Each entry contains the time of the message and its text, and
indicates whether the message was acknowledged (normal
display) or not (highlighted).
In the monitoring menu, press the key »
Scroll up and down through the log with the keys »
and »
«.
Display (example):
Exit from the menu:
Press the key »
64
« = log.
«.
«
2
1